Guidance for accurate and consistent tissue Doppler velocity measurement: comparison of echocardiographic methods using a simple vendor-independent method for local validation

Version 2 2024-03-12, 13:57
Version 1 2024-03-01, 09:27
journal contribution
posted on 2024-03-12, 13:57 authored by N. M. Dhutia, Massoud Zolgharni, C. H. Manisty, D. P. Francis, K. Willson, G. Cole, A. N. Nowbar, D. Dawson, S. Zielke, C. Whelan, J. Newton, J. Mayet

Background Variability has been described between different echo machines and different modalities when measuring tissue velocities. We assessed the consistency of tissue velocity measurements across different modalities and different manufacturers in an in vitro model and in patients. Furthermore, we present freely available software tools to repeat these evaluations.Methods and results We constructed a simple setup to generate reproducible motion and used it to compare velocities measured using three echocardiographic modalities: M-mode, speckle tracking, and tissue Doppler, with a straightforward, non-ultrasound, optical gold standard. In the clinical phase, 25 patients underwent M-mode, speckle tracking, and tissue Doppler measurements of s?, e?, and a? velocities.In vitro, the M-mode and speckle tracking velocities agreed with optical assessment. Of the three possible tissue Doppler measurement conventions (outer, middle, and inner edge) only the middle agreed with optical assessment (discrepancy ?0.20 (95% CI ?0.44 to 0.03) cm/s, P = 0.11, outer +5.19 (4.65 to 5.73) cm/s, P < 0.0001, inner ?6.26 (?6.87 to ?5.65) cm/s, P < 0.0001). A similar pattern occurred across all four studied manufacturers. M-mode was therefore chosen as the in vivo gold standard.Clinical measurements of s? velocities by speckle tracking and the middle line of the tissue Doppler showed concordance with M-mode, while the outer line overestimated significantly (+1.27(0.96 to 1.59) cm/s, P < 0.0001) and the inner line underestimated (?1.82 (?2.11 to ?1.52) cm/s, P < 0.0001).Conclusions Echocardiographic velocity measurements can be more consistent than previously suspected. The statistically modal velocity, found at the centre of the spectral pulsed wave tissue Doppler envelope, most closely represents true tissue velocity. This article includes downloadable, vendor-independent software enabling calibration of echocardiographic machines using a simple, inexpensive in vitro setup

History

School affiliated with

  • School of Computer Science (Research Outputs)

Publication Title

European Heart Journal - Cardiovascular Imaging

Volume

15

Issue

7

Pages/Article Number

817-827

Publisher

Oxford University Press (OUP) for European Society of Cardiology

ISSN

0195-668X

eISSN

1522-9645

Date Submitted

2015-11-19

Date Accepted

2014-02-14

Date of First Publication

2014-04-03

Date of Final Publication

2014-07-01

Date Document First Uploaded

2015-11-18

ePrints ID

19617

Usage metrics

    University of Lincoln (Research Outputs)

    Licence

    Exports

    RefWorks
    BibTeX
    Ref. manager
    Endnote
    DataCite
    NLM
    DC