posted on 2022-01-19, 18:24authored byJay C. Bullen, Lawrence N. Dworsky, Martijn Eikelboom, Matthieu Carriere, Alexandra Alvarez, Pascal Salaün
Increasing surface roughness of the working electrode due to the deposition of nanoparticles (a) before and (b) after continued use under acidic conditions, whereupon the charge of the reduction peak increased from 3.49 μC to 23.9 μC. (c) The increase in the intensity of the reduction peak with repeat CV measurements (including deposition at -2 V). (d) The relationship between the intensity of the reduction peak and the sensitivity of the electrode towards the detection of total As. The CV scans in clean 0.5 M H2SO4 and total As detection (167 μg L-1 As(V), 0.1 M HCl, 60 s deposition) were performed consecutively, repeatedly alternating between the two measurements.