Cutaway view of the radiometric test station.

Download (870.3 kB)
figure
posted on 2024-02-21, 18:32 authored by Paul Lebel, Susanna Elledge, Diane M. Wiener, Ilakkiyan Jeyakumar, Maíra Phelps, Axel Jacobsen, Emily Huynh, Chris Charlton, Robert Puccinelli, Prasenjit Mondal, Senjuti Saha, Cristina M. Tato, Rafael Gómez-Sjöberg

A 3D CAD model of the radiometric test setup is shown as a cross-section view. The optical path consisted of a 4f relay, imaging the tip of a 3 mm Liquid Light Guide (LLG) onto the luminometer sensor at 1× magnification. Variable numbers of neutral density filters were added or removed as required to vary the optical power, along with the lamp’s percentage output setting (0–100%). The lens relay was constructed using two 100 mm focusing lenses (Thorlabs LA1509-A), and an adjustable iris (fixed at 2 mm) to restrict the NA of the optical path. A 25 nm FWHM spectral bandpass filter (Semrock FF02-475/50-25) was employed to approximately match the wavelength of the lamp’s spectrum to that of nanoluciferase emission (460 nm). The incident optical power at the luminometer sensor was varied between 0.1 fW to nearly 1E5 fW.