MouseKidney_IMS_testdata.zip (75.76 MB)
MALDI Imaging Mass Spectrometry of a Mouse Kidney
Instrument: Autoflex III Smartbeam (Bruker, Bremen, Germany)
Linear mode (m/z 2000 - 20000)
Negative ion mode
Matrix: 2,4,6-Trihydroxyacetophenone
Spatial resolution: 150 µm
Tissue
- Mouse Kidney (BI/6, C57BL/6JRccHsd)
- Healthy
- Snap frozen in liquid N2 immediately after harvest
- 12 µm section (on Leica cryostat @ - 30°C)
Matrix application
Sublimation @ 160°C for 12 min 30 sec
Recrystallization
With 1 mL of 5% MeOH, 5% Acetic Acid @ 80°C for 2 min 30 sec (starting from RT)
Tissue wash
According to Yang and Caprioli, Anal. Chem. 2011, 83, 5728-5734 ; Rinse protocol D