Posted on 2017-05-18 - 00:00
When
quantifying endogenous plasma proteins for fundamental and
biomedical research − as well as for clinical applications
− precise, reproducible, and robust assays are required. Targeted
detection of peptides in a bottom-up strategy is the most common and
precise mass spectrometry-based quantitation approach when combined
with the use of stable isotope-labeled peptides. However, when measuring
protein in plasma, the unknown endogenous levels prevent the implementation
of the best calibration strategies, since no blank matrix is available.
Consequently, several alternative calibration strategies are employed
by different laboratories. In this study, these methods were compared
to a new approach using two different stable isotope-labeled standard
(SIS) peptide isotopologues for each endogenous peptide to be quantified,
enabling an external calibration curve as well as the quality control
samples to be prepared in pooled human plasma without interference
from endogenous peptides. This strategy improves the analytical performance
of the assay and enables the accuracy of the assay to be monitored,
which can also facilitate method development and validation.