posted on 2021-08-17, 17:23authored byJulieta S. Roldán, Alejandro Cassola, Daniela S. Castillo
The limit of detection (LOD) and the limit of quantification (LOQ) for the determination of denatured ZNS1 by icELISA with 1F5 mAb were obtained measuring the concentration of 6 different normal human sera (blank samples) and calculated by the following formulas: LOD = mean blank samples±3*SD blank samples, and LOQ = mean blank samples±10*SD blank samples. The dotted line represents the mean of the blank samples analyzed.