posted on 2021-02-03, 15:07authored byBarbora Salovska, Wenxue Li, Yi Di, Yansheng Liu
The data-independent acquisition
(DIA) performed in the latest
high-resolution, high-speed mass spectrometers offers a powerful analytical
tool for biological investigations. The DIA mass spectrometry (DIA-MS)
combined with the isotopic labeling approach holds a particular promise
for increasing the multiplexity of DIA-MS analysis, which could assist
the relative protein quantification and the proteome-wide turnover
profiling. However, the wide MS1 isolation windows employed in conventional
DIA methods lead to a limited efficiency in identifying and quantifying
isotope-labeled peptide pairs through peptide fragment ions. Here,
we optimized a high-selectivity DIA-MS named BoxCarmax that supports
the analysis of complex samples, such as those generated from Stable
isotope labeling by amino acids in cell culture (SILAC) and pulse
SILAC (pSILAC) experiments. BoxCarmax enables multiplexed acquisition
at both MS1 and MS2 levels, through the integration of BoxCar and
MSX features, as well as a gas-phase separation strategy. We found
BoxCarmax significantly improved the quantitative accuracy in SILAC
and pSILAC samples by mitigating the ratio suppression of isotope–peptide
pairs. We further applied BoxCarmax to measure protein degradation
regulation during serum starvation stress in cultured cells, revealing
valuable biological insights. Our study offered an alternative and
accurate approach for the MS analysis of protein turnover and complex
samples.