posted on 2017-12-20, 03:08 authored by Jayandiran Pillai, Ceyhan Yazicioglu, Mahad Omar, Martin G. Veller

Abstract Stent-graft migration and type I endoleaks are associated with a higher rate of reintervention and increased mortality and morbidity. This article describes a patient presented with an infrarenal aortic stent-graft which had migrated into the aortic sac with loss of all aortic neck attachment. The acutely expanding abdominal aortic aneurysm was treated by placing a second modular endograft within and above the migrated stentgraft. The patient returned 36 months later, with features of an acute myocardial infarction, severe bilateral lower limb ischemia, and renal failure. He was too ill for intervention and demised within 48 hours.