Qualification results of image-derived BSA.
figure
posted on 2018-02-14, 18:46 authored by Antoine Iannessi, Hubert Beaumont, Christophe Hebert, Claire Dittlot, Marie Noëlle Faleweea) Accuracy: Box plot of percent relative error in computing BSA of CT phantom. 14 acquisitions of the phantom were repeated with various positions. Median error was 1.54% ranging from -3.3% to 13.06%. b) Precision: Bland Altman plot of test-retest measurement of 11 thoracic and thoracic abdominal patients from the RIDER database.
History
Usage metrics
Keywords
drug dosageQuantitative radiologychemotherapy dose planningResults Precisioninter-method variabilitybody PET-CT scansBoisiBSA assessmentsevaluation Objectivebody surface areabody compositionpatient chemotherapy dosage planningreclassification rate11 CT test-retest images.Secondlychemotherapy planningDucomputer tomography-based body surface area evaluationbody CT imagesiBSA measurement