figshare
Browse
1/1
10 files

PERFORMANCE OF A PALLADIUM FILM ELECTRODE FOR THE DETERMINATION OF THALLIUM AND LEAD IN NATURAL WATER SAMPLE BY DIFFERENTIAL PULSE ANODIC STRIPPING VOLTAMMETRY

dataset
posted on 2017-11-29, 11:23 authored by Jorge Luís O. Santos, Oldair D. Leite, Clayton Ricardo Janoni, Djane S. de Jesus, Ana Maria P. dos Santos

The use of a palladium film electrode (PdFE) as an alternative electrode for the determination of thallium (Tl+) and lead (Pb2+) in water samples by anodic stripping voltammetry is proposed. Preliminary studies of the influence of the substrate were performed on the electrochemical behavior of PdFE prepared ex situ. Using a silver solid amalgam electrode as substrate (AgSAE), the effect of several parameters such as potential and deposition time, composition and pH of supporting electrolyte on the analytical response of PdFE for Tl+ by differential pulse anodic stripping voltammetry (DPASV) was initially investigated. Under best experimental conditions of deposition potential of −1.2 V and deposition time of 300s the analytical curve was linear in the Tl+ concentration from 3.0 to 100.0 µg L−1 with a relative standard deviation (RSD) <2.0%. The limit of detection (LOD) and limit of quantification (LOQ) were 0.89 µg L−1 and 2.96 µg L−1, respectively. The PdFE was successfully applied for the determination of Tl+ in water samples. Moreover, this electrode showed promising as a sensor for the simultaneous determination of Tl+ and Pb2+ and as a suitable alternative to the mercury electrode.

History

Usage metrics

    Química Nova

    Licence

    Exports

    RefWorks
    BibTeX
    Ref. manager
    Endnote
    DataCite
    NLM
    DC