Standardization of microparticle (MP) enumeration. ISTH-“Vascular Biology Workshop”. PatilRucha GhoshKanjaksha SatoskarPurnima ShettyShrimati 2013 <p>1<sup>st</sup> Graph is a graph of forward scatter (FSC-A) vs. side scatter (SSC-A) parameters showing a blend of monodisperse fluorescent beads (Megamix) of three diameters (0.5, 0.9 and 3 µm). 2<sup>nd</sup> Graph defines the MP gate which includes all events of 0.9 µm and below (microparticles) and excludes all above (cells). 3<sup>rd</sup> Graph locates the position of 3 µm beads so as to position the 10 µm counting beads 1 log ahead.</p>